This document specifies requirements and test methods for haemodialysers, haemodiafilters, haemofilters and haemoconcentrators, hereinafter collectively referred to as “the device”, for use in humans.
This document does not apply to:
— extracorporeal blood circuits;
— plasmafilters;
— haemoperfusion devices;
— vascular access devices;
— blood pumps;
— systems to prepare, maintain or monitor dialysis fluid;
— systems or equipment intended to perform haemodialysis, haemodiafiltration, haemofiltration or haemoconcentration;
— reprocessing procedures and equipment.
NOTE 1 Requirements for extracorporeal blood circuits for haemodialysers, haemodiafilters and haemofilters are specified in ISO 8637-2.
NOTE 2 Requirements for plasmafilters are specified in ISO 8637-3.