prEN IEC 61010-2-020:2024

Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-020: Particular requirements for laboratory centrifuges

Üldinfo

Kavand
Alusdokumendid
66/820/CDV; prEN IEC 61010-2-020:2024
Direktiivid või määrused
puuduvad
Standardi kavandiga saab tutvuda kommenteerimisportaalis🡭.

Standardi ajalugu

Staatus
Kuupäev
Tüüp
Nimetus
16.05.2017
Põhitekst
This clause of Part 1 is applicable except as follows:
1.1 Scope
1.1.1 Equipment included in scope
Replacement:
This part of IEC 61010 is applicable to electrically powered LABORATORY CENTRIFUGES.
It is possible that all or part of the equipment falls within the scope of one or more other Part 2 standards of IEC 61010 as well as within the scope of this standard. In that case, the requirements of those other Part 2 standards will also apply.
1.1.2 Equipment excluded from scope
Addition:
Add the following new item:
aa) IEC 60034 (Rotating electrical machinery).
1.2 Object
1.2.1 Aspects included in scope
Addition:
Add the following new items:
aa) contact with moving parts (see 7.3);
bb) LABORATORY CENTRIFUGE movement during any DISRUPTION (see 7.4.101);
cc) high energy chemical reaction after ROTOR DISRUPTION (see 7.7.2.2 l));
dd) ineffectiveness of BIOSEALS (see 13.101).

Nõutud väljad on tähistatud *

*
*
*
Standardi monitooring