prEN IEC 61010-2-034:2026

Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-034: Particular requirements for measurement equipment for insulation resistance and test equipment for electric strength

Üldinfo

Kavand
Alusdokumendid
66/875/CDV; prEN IEC 61010-2-034:2026
Direktiivid või määrused
puuduvad
Standardi kavandiga saab tutvuda kommenteerimisportaalis🡭.

Standardi ajalugu

Staatus
Kuupäev
Tüüp
Nimetus
Põhitekst
prEN IEC 61010-2-034:2022
03.05.2021
Põhitekst + muudatus
15.04.2021
Põhitekst
All Clauses and their subclauses, all Annexes, their Clauses and subclauses, and Bibliography
of IEC 61010-1:2010, IEC 61010-1:2010/AMD1:2016 and IEC 61010-1:2010/AMD2:—2 apply except as follows.
1 Scope and object
1.1 Scope
1.1.1 Equipment included in scope
Replace the existing text of 1.1.1 with the following new text:
This part of IEC 61010 specifies safety requirements for equipment for measuring insulation resistance and for equipment for testing electric strength which have an output voltage
exceeding 50 V AC or 120 V DC.
This document also applies to combined measuring equipment which has an insulation resistance measurement function or an electric strength test measurement function.
This group safety publication focusing on safety essential requirements is primarily intended to be used as a product safety standard for the products mentioned in the scope, but is also intended to be used by technical committees in the preparation of publications for products similar to those mentioned in the scope of this document, in accordance with the principles laid
down in IEC Guide 104 and lSO/lEC Guide 51.
One of the responsibilities of a technical committee is, wherever applicable, to make use of basic safety publications and/or group safety publications in the preparation of its publications.
1.1.2 Equipment excluded from scope
Add the following two new items after item h) of the list of 1.1.2:
aa) IEC 61557-8 (Electrical safety in low voltage distribution systems up to 1 000 V a.c. and 1 500 V d.c. – Equipment for testing, measuring or monitoring of protective measures – Part 8: Insulation monitoring devices for IT systems);
bb) IEC 61557-9 (Electrical safety in low voltage distribution systems up to 1 000 V AC and 1 500 V DC – Equipment for testing, measuring or monitoring of protective measures –Part 9: Equipment for insulation fault location in IT systems).
1.2.1 Aspects included in scope
Replace item c) and item h) from the list of the second paragraph of 1.2.1 with the following new item c) and item h):
c) spread of fire or arc flash from the equipment (see Clause 9);
h) HAZARDS related to the use and REASONABLY FORESEEABLE MISUSE of the equipment (see Clause 16 and Annex BB).
Insert the following new paragraph between the third paragraph and the NOTE of 1.2.1:
Annex BB and Annex CC provide guidance to equipment manufacturers on HAZARDS that should be considered for equipment intended for performing tests and measurements on hazardous conductors, including MAINS conductors and telecommunication network conductors.
1.2.2 Aspects excluded from scope
Add a new item aa) after item h) of the list of 1.2.2:
aa) HAZARDS related to the use by pupil OPERATORS in educational establishments (see IEC 61010-2-130).