Skip to main content
Back

EVS-EN ISO 10993-15:2023

Biological evaluation of medical devices - Part 15: Identification and quantification of degradation products from metals and alloys (ISO 10993-15:2019)

General information

Valid from 01.06.2023
Base Documents
ISO 10993-15:2019; EN ISO 10993-15:2023
Directives or regulations
2017/745 Medical devices (MDR)

Standard history

Status
Date
Type
Name
01.06.2023
Main
07.08.2009
Main
This document specifies general requirements for the design of tests for identifying and quantifying degradation products from final metallic medical devices or corresponding material samples finished as ready for clinical use.
This document is applicable only to those degradation products generated by chemical alteration of the final metallic device in an in vitro degradation test. Because of the nature of in vitro tests, the test results approximate the in vivo behaviour of the implant or material. The described chemical methodologies are a means to generate degradation products for further assessments.
This document is applicable to both materials designed to degrade in the body as well as materials that are not intended to degrade.
This document is not applicable to evaluation of degradation which occurs by purely mechanical processes; methodologies for the production of this type of degradation product are described in specific product standards, where available.
NOTE Purely mechanical degradation causes mostly particulate matter. Although this is excluded from the scope of this document, such degradation products can evoke a biological response and can undergo biological evaluation as described in other parts of ISO 10993.
Because of the wide range of metallic materials used in medical devices, no specific analytical techniques are identified for quantifying the degradation products. The identification of trace elements (<10–6 w/w) contained in the specific metal or alloy is not addressed in this document, nor are specific requirements for acceptable levels of degradation products provided in this document.
This document excludes the biological activity of the degradation products. (See instead the applicable clauses of ISO 10993-1 and ISO 10993-17).

Required fields are indicated with *

*
*
*
PDF
19.52 € incl tax
Paper
19.52 € incl tax
Browse standard from 2.44 € incl tax
Standard monitoring

Customers who bought this item also bought

Main

EVS-EN ISO 10993-10:2023

Biological evaluation of medical devices - Part 10: Tests for skin sensitization (ISO 10993-10:2021)
Newest version Valid from 01.03.2023
Amendment

EVS-EN ISO 10993-18:2020/A1:2023

Biological evaluation of medical devices - Part 18: Chemical characterization of medical device materials within a risk management process - Amendment 1: Determination of the uncertainty factor (ISO 10993-18:2020/Amd 1:2022)
Valid from 15.08.2023
Main

EVS-EN ISO 10993-17:2023

Biological evaluation of medical devices - Part 17: Toxicological risk assessment of medical device constituents (ISO 10993-17:2023)
Newest version Valid from 01.12.2023
Main + amendment

EVS-EN ISO 10993-18:2020+A1:2023

Biological evaluation of medical devices - Part 18: Chemical characterization of medical device materials within a risk management process (ISO 10993-18:2020 + ISO 10993-18:2020/Amd 1:2022)
Newest version Valid from 15.08.2023