Back

EVS-EN ISO 80601-2-69:2020

Medical electrical equipment - Part 2-69: Particular requirements for the basic safety and essential performance of oxygen concentrator equipment (ISO 80601-2-69:2020)

General information
Valid from 01.12.2020
Base Documents
EN ISO 80601-2-69:2020; ISO 80601-2-69:2020
Directives or regulations
None

Standard history

Status
Date
Type
Name
01.12.2020
Main
04.08.2014
Main
IEC 60601-1:2005+AMD1:2012, 1.1 is replaced by:
This document specifies requirements for the basic safety and essential performance of an oxygen
concentrator in combination with its accessories, hereafter referred to as ME equipment, intended to increase the oxygen concentration of gas intended to be delivered to a single patient. Such oxygen concentrators are typically intended for use in the home healthcare environment by a single patient in various environments including any private and public transportation as well as in commercial aircraft.
NOTE 1 Such oxygen concentrators can also be used in professional healthcare facilities.
This document is applicable to a transit-operable and non-transit-operable oxygen concentrator. This
document is applicable to an oxygen concentrator integrated into or used with other medical devices, ME equipment or ME systems.
EXAMPLE 1 An oxygen concentrator with integrated oxygen conserving equipment function or humidifier
function.
EXAMPLE 2 An oxygen concentrator used with a flowmeter stand.
EXAMPLE 3 An oxygen concentrator as part of an anaesthetic system for use in areas with limited logistical
supplies of electricity and anaesthetic gases[2].
EXAMPLE 4 An oxygen concentrator with an integrated liquid reservoir function or gas cylinder filling system
function.
This document is also applicable to those accessories intended by their manufacturer to be connected to
an oxygen concentrator, where the characteristics of those accessories can affect the basic safety or
essential performance of the oxygen concentrator.
NOTE 2 Such accessories can include, but are not limited to, masks, cannulae, extension tubing, humidifiers, carts, carrying cases, external power sources and oxygen conserving equipment.
This document does not specify requirements for oxygen concentrators for use with a medical gas pipeline system.If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.
Hazards inherent in the intended physiological function of ME equipment or ME systems within the
scope of this document are not covered by specific requirements in this document except in 7.2.13 and
8.4.1 of the general standard.
NOTE 3 See also 4.2 of the general standard.
*
*
*
PDF
28.06 € incl tax
Paper
28.06 € incl tax
Browse standard from 2.44 € incl tax
Standard monitoring

Customers who bought this item also bought

Main

EVS-EN ISO 20417:2021

Medical devices - Information to be supplied by the manufacturer (ISO 20417:2021, Corrected version 2021-12)
Newest version Valid from 17.05.2021
Main

EVS-EN ISO 80601-2-67:2020

Medical electrical equipment - Part 2-67: Particular requirements for basic safety and essential performance of oxygen-conserving equipment (ISO 80601-2-67:2020)
Newest version Valid from 01.12.2020
Main + amendment

EVS-EN 62366-1:2015+A1:2020

Medical devices - Part 1: Application of usability engineering to medical devices (IEC 62366-1:2015 + IEC 62366-1:2015/A1:2020)
Newest version Valid from 17.08.2020
Main

EVS-EN ISO 15223-1:2021

Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements (ISO 15223-1:2021)
Newest version Valid from 01.10.2021