ISO 15883-4:2018

Washer-disinfectors -- Part 4: Requirements and tests for washer-disinfectors employing chemical disinfection for thermolabile endoscopes

General information

Valid from 12.11.2018
Directives or regulations
None

Standard history

Status
Date
Type
Name
12.11.2018
Main
This document specifies the particular requirements, including performance criteria for washer-disinfectors (WD) that are intended to be used for cleaning and chemical disinfection of thermolabile endoscopes.
This document also specifies the performance requirements for the cleaning and disinfection of the washer-disinfector and its components and accessories which can be required to achieve the necessary performance criteria.
The methods, instrumentation and instructions required for type testing, works testing, validation (installation, operational and performance qualification on first installation), routine control and monitoring, and requalification of WD periodically and after essential repairs, are also specified.
NOTE 1    In addition, Annex A gives guidance on an appropriate division of responsibility for the range of activities covered by this document.
NOTE 2    WD complying with this document can also be used for cleaning and chemical disinfection of other thermolabile re-usable medical devices for which the device manufacturer has recommended and validated this method of disinfection.
WD complying with the requirements of this document are not intended for cleaning and disinfection of medical devices, including endoscopic accessories, which are heat stable and can be disinfected or sterilized by thermal methods (see ISO 15883-1:2006+Amd 1:2014, 4.1.5).
The specified performance requirements of this document do not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies.
NOTE 3    If it is considered that prion protein might be present, particular care is needed in the choice of cleaning agents and disinfectants to ensure that the chemicals used do not react with the prion protein and/or other protein in a manner that can inhibit its removal or inactivation from the load or washer-disinfector.
NOTE 4    This document can be used by prospective purchasers and manufacturers as the basis of agreement on the specification of the WD, manufacturers of endoscopes, cleaning products, and disinfecting products.

Required fields are indicated with *

*
*
*
PDF
283.20 € incl tax
Paper
283.20 € incl tax
Standard monitoring

Customers who bought this item also bought

Main

ISO 15883-1:2006

Washer-disinfectors -- Part 1: General requirements, terms and definitions and tests
Newest version Valid from 06.04.2006
Main

ISO 15883-2:2006

Washer-disinfectors -- Part 2: Requirements and tests for washer-disinfectors employing thermal disinfection for surgical instruments, anaesthetic equipment, bowls, dishes, receivers, utensils, glassware, etc.
Newest version Valid from 06.04.2006
Main

ISO 15883-3:2006

Washer-disinfectors -- Part 3: Requirements and tests for washer-disinfectors employing thermal disinfection for human waste containers
Newest version Valid from 06.04.2006
Main

ISO/TS 15883-5:2005

Washer-disinfectors -- Part 5: Test soils and methods for demonstrating cleaning efficacy
Withdrawn from 30.07.2021