Skip to main content
Tagasi
UUS

EVS-EN IEC 61267:2026

Medical diagnostic x-ray equipment - Radiation conditions for use in the determination of characteristics

Üldinfo

Kehtiv alates 16.02.2026
Alusdokumendid
IEC 61267:2025; EN IEC 61267:2026
Tegevusala (ICS grupid)
11.040.50 Radiograafiaseadmed
Direktiivid või määrused
puuduvad

Standardi ajalugu

Staatus
Kuupäev
Tüüp
Nimetus
16.02.2026
Põhitekst
08.03.2006
Põhitekst
This International Standard applies to test procedures which, for the determination of characteristics of systems or components of medical diagnostic X-ray equipment (3.2.46), require well-defined X-ray radiation condition (3.1.6).
Except for mammography, this standard does not apply to conditions where discontinuities in radiation absorption of elements are deliberately used to modify properties of the radiation beam (3.2.32) (for example by rare earth filters).
X-ray radiation condition (3.1.6) for screen-film sensitometry are not covered in this standard.
NOTE: Screen-film sensitometry is the subject of the ISO 9236 series.
This standard deals with methods for generating X-ray beams characterized by X-ray radiation conditions which can be used under test conditions typically found in test laboratories or in manufacturing facilities for the determination of characteristics of medical diagnostic X-ray equipment (3.2.46) .
Examples of such are X-ray beams emerging through the filtration from an X-ray source assembly (3.2.51) whereby the radiation field (3.2.34) includes only an insignificant amount of scattered radiation (3.2.38) . X-ray radiation condition (3.1.6) can also represent the more general case, where scattered radiation (3.2.38) emerges from an exit surface (3.1.4) of a patient (3.2.25) or a phantom (3.2.27).